Unknown error and injection stops when trying to start a run in HyStar

Unknown error and injection stops when trying to start a run in HyStar

1. Is there a sample vial in the autosampler? If not, place the sample vial. If yes, go to 2.
2. Is there a transport solvent for your uLpickup? No, place transport solvent., No, I am not using uLpickup. Go to 3.
3. RMB on Acquisition menu of HyStar and select "Commands"
4. At commands lit, LMB on +System to expand the System menu.
5. LMB on AbortBatch followed by LMB on "Execute"
6. LMB on AbortSample followed by LMB on "Execute"
7. At HyStar Acquistion page, reload sample.